Skip to content

Public FDA record

Z-0583-2020: Behind-the-Ear hearing aids, Model Numbers: Q962-DRW, T561-DRW, T562-DRW, T762-DRW, T777-DW, AMZ964-DRW,

Recall record Z-0583-2020 for GN Hearing A/S. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0583-2020
Company in FDA record
GN Hearing A/S
FDA record creation date
Nov 27, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

GN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.

Source facts

Action
The firm initiated the recall by telephone on 09/18/2019 and followed with a letter customized to each consignee. The requests the return of the affected units. The letter also requests that any entity that distributed the product notify their customers.
Initiated
2019-09-18
Root cause
Other
Terminated
2023-05-31
Distribution
Nationwide
Recall number
Z-0583-2020
Recall status
Terminated
CDRH record ID
176726
FDA record created
2019-11-27
Recall event number
83906

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗