Z-0583-2025: Spectral CT and Spectral CT Plus. Model Numbers: 728333, 728340, 728344.

Recall record Z-0583-2025 for Philips North America Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0583-2025
Company in FDA record
Philips North America Llc
FDA classification date
Dec 3, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for a plan box not updated issue during Interventional procedure with Spectral CT that may lead to a collision with the operator or with the needle placed inside the patient, and an incorrect patient ID software issue that may lead to misdiagnosis.

Source facts

Action
Philips notified consignees on about 11/16/2024, via certified mail. Consignees were instructed to identify any affected units on hand, identify if those units contain the affected software version, complete and return the response form, and circulate the Urgent Field Safety Notice Letter to all users of the device. To protect the user and patient from the issue of Plan Box not Updated, remain vigilant and confirm table position during procedures. Press any of the Emergency STOP buttons to stop unintended patient table motion and follow safety instructions in Section 6 of the IFU: Working with Specialized Exams. To avoid the issue of Wrong Patient ID with Anonymous patient, the operator should verify the patient information loaded into the patient data field is correct before scanning and rename the Anonymous record with the patient s name at the time of the scan.
Initiated
2024-11-16
Root cause
Under Investigation by firm
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, FL, GA, HI, IN, KY, MA, MD, MN, NY, OH, OR, PA, TX, VT, WA, WV and the countries of Australia, Austria, Belgium, China, Czech Republic, Denmark, Ecuador, France, Germany, Hong Kong, Israel, Italy, Japan, Jordan, Macao, Malaysia, Netherlands, New Zealand, Norway, Palestine, Panama, Poland, Portugal, Singapore, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Vietnam.
Recall number
Z-0583-2025
Recall status
Open, Classified
CDRH record ID
211267
FDA classified
2024-12-03
Recall event number
95810

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗