Skip to content

Public FDA record

Z-0583-2026: DermaSensor. Model/Catalog Number: 10101. Software Version: 3.0.2. Software-aided adjunctive diagnostic

Recall record Z-0583-2026 for DERMASENSOR INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0583-2026
Company in FDA record
DERMASENSOR INC
FDA classification date
Nov 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for device to not meet specifications resulting in incorrect results or delay of referral for patients.

Source facts

Action
DermaSensor began notification of consignees on about 10/13/2025 via telephone call with a follow up letter. The telephone call notified consignees of the issue and confirmed if the affected units were or were not on hand. The letter then sent to the consignees again advised them of the issue and instructed them to examine inventory for affected units, to complete and return the Customer Notification Response Form, and arrange a visit with DermaSensor for the device correction. Once the affected unit is examined, DermaSensor can determine if the device contains the component issue and if it does, DermaSensor will provide the consignee with the dates the scans were taken and the correct results. If the consignee's device does not contain the component issue, no further actions need to be taken.
Initiated
2025-10-13
Root cause
Process control
Distribution
US distribution to MD, FL, AZ, MS, PA, NY, CT, CA.
Recall number
Z-0583-2026
Recall status
Open, Classified
CDRH record ID
216559
FDA classified
2025-11-21
Recall event number
97780

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗