Open original FDA record ↗Recall record
- FDA identifier
- Z-0583-2026
- Company in FDA record
- DERMASENSOR INC
- FDA classification date
- Nov 21, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential for device to not meet specifications resulting in incorrect results or delay of referral for patients.
Source facts
- Action
- DermaSensor began notification of consignees on about 10/13/2025 via telephone call with a follow up letter. The telephone call notified consignees of the issue and confirmed if the affected units were or were not on hand. The letter then sent to the consignees again advised them of the issue and instructed them to examine inventory for affected units, to complete and return the Customer Notification Response Form, and arrange a visit with DermaSensor for the device correction. Once the affected unit is examined, DermaSensor can determine if the device contains the component issue and if it does, DermaSensor will provide the consignee with the dates the scans were taken and the correct results. If the consignee's device does not contain the component issue, no further actions need to be taken.
- Initiated
- 2025-10-13
- Root cause
- Process control
- Distribution
- US distribution to MD, FL, AZ, MS, PA, NY, CT, CA.
- Recall number
- Z-0583-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216559
- FDA classified
- 2025-11-21
- Recall event number
- 97780