Reason stated in the source
Some of the hand controllers of the Symbia S & T systems are missing a resistor switch, and the absence of the switch may lead to unintended system motion during camera set up activities.
Source facts
- Action
- Siemens issued an Urgent Field Correction Recall letters dated November 18, 2009 and December 11, 2009 to the affected users of the Symbia S and Symbia T Systems, respectivley. The customers were advised of precautionary measures and that Siemens will replace the affected device. They also were advised that a Siemens Customer Service Representative would contact them within a few days to schedule a replacement of the device. For further information, contact Siemens Medical Solutions USA, Inc. at 1-800-888-7436 (USA).
- Initiated
- 2009-12-11
- Root cause
- Process control
- Terminated
- 2012-08-27
- Distribution
- Worldwide Distribution -- United States, Algeria, Australia, Austria, Belgium, Canada, Chile, China, Finland, France, Germany, Hong Kong, India, Indonesia, Italy, Korea, Mexico, Netherlands, Norway, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Taiwan, Turkey, Venezuela and Vietnam.
- Recall number
- Z-0584-2010
- Recall status
- Terminated
- CDRH record ID
- 86868
- FDA classified
- 2009-12-29
- Recall event number
- 53919