- Action
- Stryker Instruments initiated a voluntary recall of the RF MultiGen, and Refurb RF MultiGen because the instructions for use contained incorrect instructions on sterilizing the device. Stryker distributed customer notices via certified mail on 10/11/2016.
Customer's were instructed to do the following: Immediately review this recall notification.
2. Enclosed was a revised version of the IVS Care Instructions. Please distribute the revised version to the appropriate person or department.
3. Remove and discard any previously distributed IVS Care Instructions or Instructions for Use.
4. Please complete, sign, and return the enclosed Business Reply Form (BRF) acknowledging your receipt and understanding of this notification.
5. Fax the completed BRF to Stryker Instruments Regulatory Department, 866-521-2762, or scan and email a copy to kellyjo.davis@stryker.com. Customers with questions can contact: 269 323 7700 .
- Initiated
- 2016-10-11
- Root cause
- Labeling Change Control
- Terminated
- 2017-07-25
- Distribution
- Domestic:AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY
Foreign:France, Germany, Canada, UK, Lebanon, South Africa, Chile, Hong Kong, Poland, Argentina, Spain, Greece, Italy, India, Brazil, Switzerland, Netherlands, Sweden
VA/DOD:
- Recall number
- Z-0584-2017
- Recall status
- Terminated
- CDRH record ID
- 150479
- FDA classified
- 2016-11-16
- Recall event number
- 75485