Z-0584-2026: The LeadCare II Blood Lead Test System relies on electrochemistry and a unique Sensor to detect lead in w

Recall record Z-0584-2026 for Magellan Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0584-2026
Company in FDA record
Magellan Diagnostics, Inc.
FDA classification date
Nov 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Calibration Buttons being used in the kitting process were incorrectly labeled as Lot 2323M for lot 2333M. The use of mislabeled calibration Button with Lot #2333 can lead to delayed or erroneous high or erroneous low results.

Source facts

Action
On March 3, 2024, MEDICAL DEVICE FIELD CORRECTION letters were sent to customers. Immediate Actions: 1. Identify existing inventory of impacted product and suspend its distribution. 2. Identify customers that received the impacted product and notify them about this field action. 3. Return remaining impacted product to Magellan Diagnostics, Inc. for destruction. To initiate the return process, ensure that you note the quantity of product to be returned on your Confirmation of Notification Form. Actions to Be Taken by Magellan: 1. Magellan Diagnostics will continue to investigate the cause of the labeling error and initiate appropriate preventive actions. 2. Magellan will replace unexpired impacted product upon request. Product will be replaced based on availability, and on a first-come first-served basis.
Initiated
2024-02-23
Root cause
Labeling mix-ups
Distribution
US Nationwide distribution.
Recall number
Z-0584-2026
Recall status
Open, Classified
CDRH record ID
216773
FDA classified
2025-11-21
Recall event number
97858

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗