- FDA identifier
- Z-0585-2010
- Company in FDA record
- Teleflex Medical
- FDA classification date
- Dec 29, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Sensor causes the oxygen monitor to display prematurely a Low Sensor warning which indicates an early maintenance requirement for the instrument. This may result in interrupted treatment.
Source facts
- Action
- Consignees were notified via next day delivery on or about 11/25/2009. Each customer received a letter requesting acknowledgement of the recall and a request to return the affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. A second and third mailing to non-responding customers on the consignee list will be made, or as an alternative, the third contact may be by phone, e-mail, or fax. Affected products received by Teleflex Medical will be destroyed or reworked and replacement product or credit will be provided to the consignee. Records of contacts and of destruction or rework will be maintained. Questions are directed to the company at 1-866-246-6990.
- Initiated
- 2009-11-25
- Root cause
- Vendor change control
- Terminated
- 2011-04-07
- Distribution
- FL, KY, MI, OH, PA, SD, TN, Australia and Canada.
- Recall number
- Z-0585-2010
- Recall status
- Terminated
- CDRH record ID
- 86890
- FDA classified
- 2009-12-29
- Recall event number
- 53937