INTERNAL SHORT CIRCUIT - MAY ALLOW FOR STAND AND TABLE TO MOVE SPONTANEOUSLY IN UPWARD OR DOWNWARD MOVEMENT
Source facts
Action
The firm issued a letter dated January 10, 2003 to their customers advising of the problem and that a respresentative would be contacting customers to arrange a time to visit to install new interface board.
This is being conducted under FCO 70800007
Initiated
2003-01-10
Root cause
Other
Terminated
2005-04-11
Distribution
The firm distributed to 22 hospitals/medical centers located in CA, FL, IA, IL, IN, KY, MN, MO, NC, ND, NY, OH, SC, TN, TX, VA, and WV.
Recall number
Z-0586-03
Recall status
Terminated
CDRH record ID
25769
FDA classified
2003-02-27
Recall event number
25406