Open original FDA record ↗Recall record
- FDA identifier
- Z-0586-06
- Company in FDA record
- B. Braun Medical Inc
- FDA classification date
- Mar 3, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
B. Braun Medical has been notified by the manufacturer of the product (Stedim) that a small percentage of the lots leak from the set port when the set port is opened. The leak was identified by Stedim to be due to the incorrect placement of a septum in the set port. The set closure is intact and the leak remains internal until the set port is opened.
Source facts
- Action
- A recall notification letter was sent to the 22 customers who received the product between 12/1/05 and 1/11/06. The recall packages included a letter addressed to either the Director of Pharmacy/Director of Nursing, Materials Manager, or Distributor, as appropriate. They were mailed via US Post Office certified return receipt mail. Customers were instructed by letter to return all product at their facility to B. Braun Medical, Inc. in Breinigsville Pennsylvania. A Recall Information Form and a postage paid B. Braun Medical, Inc. return envelope were included in the recall package for the customer''s use.
- Initiated
- 2006-02-06
- Root cause
- Other
- Terminated
- 2008-05-28
- Distribution
- Nationwide to: CA, CT, FL, GA, KS, MD, MI, NC, NJ, NY, SC, TN, & TX
- Recall number
- Z-0586-06
- Recall status
- Terminated
- CDRH record ID
- 44161
- FDA classified
- 2006-03-03
- Recall event number
- 34598