Skip to content

Public FDA record

Z-0586-06: B. Braun Hyperformer EVA Mixing Container with Hyperformer Compatible Connector, 4000ml size, Catalog Num

Recall record Z-0586-06 for B. Braun Medical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0586-06
Company in FDA record
B. Braun Medical Inc
FDA classification date
Mar 3, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

B. Braun Medical has been notified by the manufacturer of the product (Stedim) that a small percentage of the lots leak from the set port when the set port is opened. The leak was identified by Stedim to be due to the incorrect placement of a septum in the set port. The set closure is intact and the leak remains internal until the set port is opened.

Source facts

Action
A recall notification letter was sent to the 22 customers who received the product between 12/1/05 and 1/11/06. The recall packages included a letter addressed to either the Director of Pharmacy/Director of Nursing, Materials Manager, or Distributor, as appropriate. They were mailed via US Post Office certified return receipt mail. Customers were instructed by letter to return all product at their facility to B. Braun Medical, Inc. in Breinigsville Pennsylvania. A Recall Information Form and a postage paid B. Braun Medical, Inc. return envelope were included in the recall package for the customer''s use.
Initiated
2006-02-06
Root cause
Other
Terminated
2008-05-28
Distribution
Nationwide to: CA, CT, FL, GA, KS, MD, MI, NC, NJ, NY, SC, TN, & TX
Recall number
Z-0586-06
Recall status
Terminated
CDRH record ID
44161
FDA classified
2006-03-03
Recall event number
34598

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗