Open original FDA record ↗Recall record
- FDA identifier
- Z-0587-2015
- Company in FDA record
- Heart Sync, Inc.
- FDA classification date
- Dec 24, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the
Source facts
- Action
- The firm, HeartSync, sent an "Voluntary Field Safety Alert for Heart Sync Multi-function Defibrillator Electrodes" notice dated 11/11/2014 to customers via email notifying them of the incompatibility with the Philips FR3 and FRX AEDs and requesting response to the notice. On 12/1/2014 (dated 11/26/2014) Heart Sync issued a Voluntary Device Correction via Press Release. The Press Release included a description of the reason for the recall, affected product, and instruction that the Philips FR3 and FRx AED units should only be used with electrodes specified in the (Philips) equipment manuals. For further information or to report a problem, please contact Heart Sync at 734-213-5530, 24 hours a day, 7 days a week, ,or email at Jahana@heartsync.net.
- Initiated
- 2014-11-11
- Root cause
- Device Design
- Terminated
- 2015-11-24
- Distribution
- Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.
- Recall number
- Z-0587-2015
- Recall status
- Terminated
- CDRH record ID
- 132028
- FDA classified
- 2014-12-24
- Recall event number
- 69875