Skip to content

Public FDA record

Z-0588-2010: Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB,

Recall record Z-0588-2010 for MAQUET Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0588-2010
Company in FDA record
MAQUET Inc.
FDA classification date
Jan 5, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.

Source facts

Action
Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
Initiated
2009-11-17
Root cause
Device Design
Terminated
2012-06-04
Distribution
Worldwide distribution
Recall number
Z-0588-2010
Recall status
Terminated
CDRH record ID
86656
FDA classified
2010-01-05
Recall event number
53843

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗