Software defect that causes an incorrect dose compensation function to be applied to the internal dose ionization chamber when either pressure or temperature sensor malfunction. This could result in an incorrect dose delivery of no more than 5%.
Source facts
Action
Mevion Medical Systems sent an Important Safety Notice letter to the affected customer. The letter identified the affected product, problem and actions to be taken. A software update addressing the issue will be provided to the customer on or about December 5, 2014.
For questions or concerns, email lbouchet@mevion.com
or call 978-540-1555.
Initiated
2014-11-10
Root cause
Software design
Terminated
2015-03-18
Distribution
US Distribution in the state of MO
Recall number
Z-0588-2015
Recall status
Terminated
CDRH record ID
131530
FDA classified
2014-12-12
Recall event number
69778