Skip to content

Public FDA record

Z-0588-2026: Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table

Recall record Z-0588-2026 for Pivotal Health Solutions, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0588-2026
Company in FDA record
Pivotal Health Solutions, Inc.
FDA classification date
Nov 24, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.

Source facts

Action
Pivotal Health Solutions issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 10/9/2025 via certified mail and Fed Ex. The notice explained the issue with the device, risk to health and requested the following: Immediately stop using table(s) with heat. The heat function must be disabled. You have been identified as having been sold a Quantum Table with heat. The serial numbers of the tables impacted are below. If you are unsure if your table has a heat function, please identify the serial number on the Quantum table(s), located on the lower section of table by the power cable inlet, and verify the serial number is on the list. The notice also supplied instructions for disconnecting and disabling the heat function o the Q400 tables. Contact information for questions: o Matt White (matt.white@pivotalhealthsolutions.com) or Julee Driver (julee.driver@pivotalhealthsolutions.com), Quality Dept, at 605-753-0110 o Brenda Anderson (brenda.anderson@pivotalhealthsolutions.com), Service Dept at 605-753-0110
Initiated
2025-10-09
Root cause
Device Design
Distribution
Distribution to US nationwide, Canada and Japan.
Recall number
Z-0588-2026
Recall status
Open, Classified
CDRH record ID
216790
FDA classified
2025-11-24
Recall event number
97861

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗