Reason stated in the source

Pivot joint on these ceiling mount lights may fail and lights may fall onto patients.

Source facts

Action
Firm will ask for subdistribution information from thier distributors via letter on 1/30/2004. Letters will be sent directly to users. Firm will retrofit all sites with suspect units with new pivot assemblies.
Initiated
2004-02-19
Root cause
Component design/selection
Terminated
2012-01-25
Distribution
Nationwide.
Recall number
Z-0589-04
Recall status
Terminated
CDRH record ID
31412
FDA classified
2004-07-20
Recall event number
28229

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗