Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
Source facts
Action
Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
Initiated
2009-11-17
Root cause
Device Design
Terminated
2012-06-04
Distribution
Worldwide distribution
Recall number
Z-0589-2010
Recall status
Terminated
CDRH record ID
86682
FDA classified
2010-01-05
Recall event number
53843