Open original FDA record ↗Recall record
- FDA identifier
- Z-0589-2011
- Company in FDA record
- Cardiac Science Corporation
- FDA classification date
- Dec 10, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some units may be non-operational during a rescue, due to a defective PCB interconnect on traces connected to the reed relay. This will result in an inability to analyze the patient's heart rhythm or deliver therapy during a rescue.
Source facts
- Action
- Cardiac Science issued a Urgent Medical Device Corrective Action letter dated April 28, 2010 requesting GE Healthcare cease distribution of affected devices and to instruct end users of distributed devices to schedule a service call to have the device corrected. Cardiac will send rework kits no later than June 30, 2010. CSC can be contacted at 44 161 926 0011 to arrange return material authorization and replacement of affected devices.
- Initiated
- 2010-04-28
- Root cause
- Nonconforming Material/Component
- Terminated
- 2011-12-08
- Distribution
- Worldwide Distribution: Throughout the US, and in the countries of Germany, Korea, Peru, Singapore, and Spain.
- Recall number
- Z-0589-2011
- Recall status
- Terminated
- CDRH record ID
- 92350
- FDA classified
- 2010-12-10
- Recall event number
- 55977