Open original FDA record ↗Recall record
- FDA identifier
- Z-0589-2021
- Company in FDA record
- RAYSEARCH LABORATORIES AB
- FDA record creation date
- Dec 15, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.
Source facts
- Action
- On November 11, 2020, the firm issued Field Safety Notices to affected customers via email. Customers were informed of the affected versions and the product issue. ACTIONS TO BE TAKEN BY THE USER: " Always verify in the RayStation/RayPlan GUI that all relevant regions of interest (ROIs) are within the dose grid. " If using the treatment plan report for evaluation of clinical goal results, verify the percentage outside the dose grid (% outside grid) for each ROI in the RayStation/RayPlan GUI. " In the treatment plan report, be aware that the Couch pitch angle and Couch roll angle values for setup beams may be incorrect. " Use exported DICOM plans, not the treatment plan report, for transferring treatment parameters to the delivery system. " Please educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. This issue will be resolved in the next version of RayStation and RayPlan, scheduled for market release in November 2020 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation and RayPlan affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.
- Initiated
- 2020-11-11
- Root cause
- Software design
- Terminated
- 2023-05-09
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0589-2021
- Recall status
- Terminated
- CDRH record ID
- 184630
- FDA record created
- 2020-12-15
- Recall event number
- 86832