Z-0589-2021: RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all serv

Recall record Z-0589-2021 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0589-2021
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Dec 15, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.

Source facts

Action
On November 11, 2020, the firm issued Field Safety Notices to affected customers via email. Customers were informed of the affected versions and the product issue. ACTIONS TO BE TAKEN BY THE USER: " Always verify in the RayStation/RayPlan GUI that all relevant regions of interest (ROIs) are within the dose grid. " If using the treatment plan report for evaluation of clinical goal results, verify the percentage outside the dose grid (% outside grid) for each ROI in the RayStation/RayPlan GUI. " In the treatment plan report, be aware that the Couch pitch angle and Couch roll angle values for setup beams may be incorrect. " Use exported DICOM plans, not the treatment plan report, for transferring treatment parameters to the delivery system. " Please educate planning staff and all users about this workaround. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. This issue will be resolved in the next version of RayStation and RayPlan, scheduled for market release in November 2020 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation and RayPlan affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.
Initiated
2020-11-11
Root cause
Software design
Terminated
2023-05-09
Distribution
US Nationwide distribution.
Recall number
Z-0589-2021
Recall status
Terminated
CDRH record ID
184630
FDA record created
2020-12-15
Recall event number
86832

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗