Open original FDA record ↗Recall record
- FDA identifier
- Z-0589-2026
- Company in FDA record
- Argon Medical Devices, Inc
- FDA classification date
- Nov 24, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to complaints of increased resistance when advancing the dilator within the introducer sheath. This issue could potential result in procedural delays and/or risk of venous intimal injuries if excessive force is applied
Source facts
- Action
- On 09/12/2025, the firm sent via certified mail an "Urgent - Product Recall Notice" to customers. On 9/2025 the firm sent a follow-up "URGENT - MEDICAL DEVICE RECALL NOTICE" dated 9/29/2025 informing them that Argon Medical Devices is voluntarily recalling five affected lots of the Option ELITE Vena Cava Filter System products in response to reports of increased resistance when advancing the dilator within the introducer sheath. Customers are instructed to: 1. Immediately discontinue use of the affected devices listed in the Recall Letter. 2. Segregate and quarantine the affected devices. 3. Complete and return the response form included at the end of this notification via regular mail or email by 9/26/2025. 4. Return all affected devices to Argon Medical Devices using the instructions provided below. 5. Share this recall notice with the following parties to ensure consignees receive a copy and respond accordingly: a. Anyone within their organization who may use this device. b. Any organization where the potentially affected products have been transferred or further distributed. c. Anyone who needs to be aware within their organization. The product may be returned at Argon s expense to the address below with the returned goods authorization (RGA# 29074) clearly marked. RGA# 29074 Argon Medical Devices, Inc. 1445 Flat Creek Road Athens, TX 75751 USA For any questions contact/email Clayton.Jones@argonmedical.com, or Scott.Bishop@argonmedical.com.
- Initiated
- 2025-09-12
- Root cause
- Equipment maintenance
- Distribution
- U.S. (nationwide) distribution to states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, and WV. O.U.S. (international) to countries of: China, Panama and Trinidad and Tobago
- Recall number
- Z-0589-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216137
- FDA classified
- 2025-11-24
- Recall event number
- 97626