There is an issue regarding poorly processed and/or damaged biopsy tissue specimens on the HistoCore PEGASUS / HistoCore PEGASUS Plus resulting from an incorrect assignment of the carryover value set for created or edited protocols.
Source facts
Action
Leica issued an URGENT Medical Device Recall notice to its consignees on 11/20/2023 via email. The notice explained the issue with the device and instructed the consignee to ensure the selection of the correct setting of the carryover value for al created or edited protocols and follow the instructions in the IFU.
Initiated
2023-11-20
Root cause
Other
Distribution
US, Nationwide
Recall number
Z-0590-2024
Recall status
Open, Classified
CDRH record ID
204461
FDA classified
2023-12-20
Recall event number
93467