Skip to content

Public FDA record

Z-0591-2010: Bausch & Lomb SoFlex SE Foldable Intraocular Lens. UPC 4001404, Rx Only, STERILE. Manufactured by: Bausc

Recall record Z-0591-2010 for Bausch & Lomb Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0591-2010
Company in FDA record
Bausch & Lomb Inc
FDA classification date
Jan 11, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Bausch & Lomb Incorporated, Rochester, NY is recalling their Silicone L161SE Intraocular Lens Lot 4916928 (23.5D) after receiving three customer complaints for what appears to be opacification of the intraocular lens. The opacification occurred within one week of implantation. In each reported case the surgeon explanted the lens.

Source facts

Action
Bausch & Lomb, Inc. notified consignees via phone beginning November 19, 2009 and by follow-up letter dated November 23, 2009. Titled "Urgent - Medical Device Recall", the letter instructed users to discontinue implantation of the affected product and return any unused lenses along with the enclosed acknowledgement form to the firm. For further information, contact Bausch & Lomb at 1-585-338-6612.
Initiated
2009-11-19
Root cause
Other
Terminated
2010-07-22
Distribution
Lenses were distributed to one distributor and 35 retailers. Worldwide Distribution -- Distributed to Canada, Taiwan and throughout the United States.
Recall number
Z-0591-2010
Recall status
Terminated
CDRH record ID
86910
FDA classified
2010-01-11
Recall event number
53949

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset: