Open original FDA record ↗Recall record
- FDA identifier
- Z-0592-2024
- Company in FDA record
- Busse Hospital Disposables, Inc.
- FDA classification date
- Jan 3, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Firm is recalling all kits containing recalled lots of Nurse Assist Sterile 0.9% Normal Saline. Nurse Assist identified product in distribution for which the required sterility assurance level (SAL) of 10^-6 cannot be guaranteed. Risks to patients if product is used include blood stream infections, urinary tract infections, or respiratory infections; in the worst-case scenario, there is risk of sepsis or death.
Source facts
- Action
- On November 21, 2023, the firm notified affected distributors via URGENT MEDICAL DEVICE RECALL letters. Customers were asked to segregate and destroy any affected kits. Distributors should also inform their customers of the recall. Customers should return any affected product to their distributor. For questions, email NurseAssistRecall@busseinc.com
- Initiated
- 2023-11-30
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide distribution.
- Recall number
- Z-0592-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204700
- FDA classified
- 2024-01-03
- Recall event number
- 93546