Z-0593-2010: Powerlink System with IntuiTrak Delivery System (Endovascular Graft): Model Number 34-34-80L, LOTS W09-2

Recall record Z-0593-2010 for Endologix Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0593-2010
Company in FDA record
Endologix Inc
FDA classification date
Jan 11, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

This action is being taken because the product has the potential for the polyimide tubing to detach from the inner core due to the bond failure. If this occurs, treatment may be delayed or interrupted. The firm's initial health hazard analysis indicated the event may cause temporary or medically reversible adverse health consequences or an outcome where the probability of serious adverse health c

Source facts

Action
The recall communication was initiated on 11/25/09 by phone call and followed by recall letter via Fed-Ex Overnight. The recall letter informed the consignees of the reason for recall and products affected. Users were instructed to share this information with physicians that performed the affected procedures at their facility. Endologix informed their consignees that their personnel would retrieve the affected devices and return it to Endologix at no cost and would replace the affected units. Consignees with questions or comments about the notice are directed to (800) 983-2284.
Initiated
2009-11-25
Root cause
Process design
Terminated
2010-05-07
Distribution
Recalled devices were distributed to medical centers and hospitals in OR, PA, FL, TX, MI and LA.
Recall number
Z-0593-2010
Recall status
Terminated
CDRH record ID
86988
FDA classified
2010-01-11
Recall event number
53983

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗