Open original FDA record ↗Recall record
- FDA identifier
- Z-0593-2014
- Company in FDA record
- Philips Medical Systems, Inc.
- FDA classification date
- Dec 27, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Error in the printing of the therapy energy setting labels on Philips HeartStart M3536A MRx Monitor/Defibrillators. The first energy setting is labeled 10 when it should be labeled1-10
Source facts
- Action
- Philips sent a Customer Information Letter dated June 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. During an EI of the firm, the firm had not reported the correction to FDA. The Customer Information letter informs the customers what the behavior is and under what circumstances. The correction consisted of a replacement label that was provided free of charge to all units affected by this issue. Philips Healthcare representatives have been contact customers with affected devices to arrange for installation of the replacement label. To correct this issue, Philips will replace the label free of charge. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the replacement label. This issue does not affect current product shipments. I f you need any further information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377. Philips apologizes for any inconvenience caused by this problem.
- Initiated
- 2013-06-07
- Root cause
- Labeling Change Control
- Terminated
- 2017-03-22
- Distribution
- Worldwide Distribution - US Distribution including the states of IN, MS, TX and VA., and the countries of DENMARK, FRANCE, GERMANY, SPAIN and SWITZERLAND.
- Recall number
- Z-0593-2014
- Recall status
- Terminated
- CDRH record ID
- 124617
- FDA classified
- 2013-12-27
- Recall event number
- 67052