Z-0594-2019: Cascade Abrazo c-ACT-LR Test Card IVD FOR IN VITRO DIAGNOSTIC USE Store at 2 to 8 C HELENA LABORATORIES 0

Recall record Z-0594-2019 for Helena Laboratories, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0594-2019
Company in FDA record
Helena Laboratories, Inc.
FDA record creation date
Dec 10, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to some cards in this lot experiencing variations in wave form, resulting in some cards with shorter or longer times than expected

Source facts

Action
On 04/20/2016 the firm, Helena Laboratories, contacted affected users via email, and then sent a "WITHDRAWAL NOTIFICATION" (dated April 25, 2016) on 04/27/2016. The notification described the product, problem and actions to be taken. The firm instructed the customers/distributors to disposes of affected lot, as a preventive quality measure. The Withdrawal Notification for Removal of Product from Inventory stated, This lot is to be handled by Distributors, as follows: ___ Prevent any shipments or scheduled demonstrations with this lot ___ Remove this lot from Available Inventory and discard the cards ___ Place cards in Quarantine pending disposal ___ Contact International Sales for credit and delivery of replacement cards Verify amount REMOVED FROM STOCK: Total number pulled: ____________________ Date _____________________________ The following signifies written certification that this lot has not been sent to an end user: Verified by: _____________________________________________________________ Please complete this form and return a copy to Helena Laboratories, via: FAX: 409-842-1874 or e-mail to jjones@helena.com For questions, contact Mgr., Technical Services at: 1 (409) 842-3714 ext. 1177. *Reason for removal: Some cards in this lot are experiencing variations in wave form, resulting in possible shorter, or longer, times than expected. While this issue, if present, should be found via use of the Quality Control materials, due to its intermittent nature in the cards, this alone cannot be relied upon.
Initiated
2016-04-20
Root cause
Device Design
Terminated
2024-02-15
Distribution
US Distribution to state of TX and internationally to: Australia, Netherlands, Italy, and England.
Recall number
Z-0594-2019
Recall status
Terminated
CDRH record ID
168364
FDA record created
2018-12-10
Recall event number
81356

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗