Z-0595-2024: CARDIOHELP-i Heart Lung Support System: Blood oxygenation and carbon dioxide removal system used to pump

Recall record Z-0595-2024 for Maquet Medical Systems USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0595-2024
Company in FDA record
Maquet Medical Systems USA
FDA classification date
Sep 30, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

CARDIOHELP-i System was not properly tested to measure leakage current

Source facts

Action
Maquet Cardiopulmonary GmbH (MCP)/Getinge issued Urgent Medical Device Correction Letter to U.S, consignees November 14, 2023 via FedEx. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any affected CARDIOHELP-i in your inventory. " Whether or not your facility has affected product(s) listed in this notice, please complete and sign the attached MEDICAL DEVICE - CORRECTION RESPONSE FORM (Page 4) to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to Cardiopulmonary-cardiohelpleakage2023.cp@getinge.com or by faxing the form to 1-800-530-6084. " If you report having any affected CARDIOHELP-i devices, a local Getinge representative will contact you to arrange a system check of the device to ensure that leakage current is within specifications. " Affected CARDIOHELP-i devices are not requested to be returned and can be used as usual. " Please forward this information to all current and potential CARDIOHELP-i users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Type of Action by Getinge: Getinge has identified the issue to be related to a supplier using an incorrect drawing to produce a tool used to measure leakage current of the CARDIOHELP-i. This issue has been corrected by the supplier. MCP/Getinge will contact you to arrange for a MCP/Getinge service technician to perform a system check of any affected device you report to confirm leakage current is within specifications (IEC 62353). If a system fails the performed test, the MCP/Getinge service technician will replace the system s sensor panel for you. MCP/Getinge service may have already performed the required system check of your device during a past preventative maintenance. MCP/Getinge will check service records to confirm. MCP
Initiated
2023-11-14
Root cause
Process design
Distribution
Nationwide-WW Distribution to Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, French Polynesia, Germany, Ghana, Greece, Hong Kong, Iceland, India, Iran, Islamic Republic Of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic Of Kuwait, Lebanon, Luxembourg, Macao, Maldives, Martinique, Mayotte, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom,Vietnam, Yemen
Recall number
Z-0595-2024
Recall status
Open, Classified
CDRH record ID
204422
FDA classified
2024-09-30
Recall event number
93462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗