Z-0596-2016: Philips Lumify Diagnostic Ultrasound, Catalogue Number: 795216 Part Number: 989605449841 with 45356184533

Recall record Z-0596-2016 for Philips Ultrasound, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0596-2016
Company in FDA record
Philips Ultrasound, Inc.
FDA classification date
Jan 6, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Color Flow direction is displayed incorrectly in Lumify 1.0. The system displayed Color Flow direction does not correctly represent the annotated Color Bar or Velocity Markers. (e.g.) When the Color Bar conveys that Red is to be displayed for color flow toward the Transducer, the system displays Blue for color flow toward the Transducer. This could result in misdiagnosis in some studies.

Source facts

Action
Philips began mailing the URGENT - Medical Device Correction Philips Model Lumify Ultrasound System, FSN 79500372, dated 2015 DEC 11, to consignees. Consignees are instructed to do the following: 1. Check the version of software on your system using the instructions provided above on How to Identify Affected Products. Philips has corrected this issue in version 1.0.1 and higher. 2. Connect the tablet to the internet. 3. The system will alert you to download the upgrade from Google Play. 4. Follow all on screen to upgrade the software version. If consignees cannot promptly connect to the internet to download the upgraded software, Philips recommends that you do not use Color Flow. In accordance with the subscription terms and conditions, you will be required to reconnect the system to the internet within no more than 30 days. If consignees do not connect their device to the internet after 30 days, they will be unable to use the system. Consignees with questions should contact your local Philips representative or Philips Customer Service at 1-800-722-9377.
Initiated
2015-12-09
Root cause
Software design
Terminated
2016-08-08
Distribution
distributed in CA, CT, ND, NV, OR, RI, TN, and WA.
Recall number
Z-0596-2016
Recall status
Terminated
CDRH record ID
142399
FDA classified
2016-01-06
Recall event number
72879

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗