Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs, depletion can take place within a few hours to a few days, after which there is a complete loss of device function.
Source facts
Action
Medtronic Sales Representatives contacted affected physicians starting 02/10/05 to deliver an Important Patient Management Information. The letter recommended patient managment options including quarterly follow-ups, programming Patient Alert to 'On-High', provide magnet to patients to check device status and inform patients that should they expereicnce warmth in the area surrounding the ICD to seek prompt care.
Initiated
2005-02-09
Root cause
Other
Terminated
2006-12-28
Distribution
worldwide
Recall number
Z-0597-05
Recall status
Terminated
CDRH record ID
37239
FDA classified
2005-03-11
Recall event number
31142