Open original FDA record ↗Recall record
- FDA identifier
- Z-0597-2016
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Jan 6, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential post-processing software issue when using tabcard "4D" on X-Workplace with software version VD10E. A too small measurement in the MPRs of a volume when images are acquired with a CT scanner with a tilted gantry. In a 2x2 layout, the 4D tabcard will show an incorrect, too small length measurement. This could result in selecting a device of the wrong size, which then needs to be exchanged.
Source facts
- Action
- Siemens sent an Important Customer Safety Notice dated November 17, 2015. The Notice identified the product, the problem, and the action to be taken by the customer. The potential issue will be remedied with a software update via Update Instructions AX067/15/S. Siemens service organization will contact customers to arrange a date for the installation of the software update. Customers with questions were advised to call 800-888-7436.
- Initiated
- 2015-11-18
- Root cause
- Software design
- Terminated
- 2016-12-05
- Distribution
- Nationwide Distribution including: AL, AZ, CA, CO, FL, IA, ID, IL, IN, MA, MD, MI, MN, MO, MT, NC, NY, OH, OK, PA, SC, TN, TX, VA, and WI.
- Recall number
- Z-0597-2016
- Recall status
- Terminated
- CDRH record ID
- 142124
- FDA classified
- 2016-01-06
- Recall event number
- 72751