Z-0597-2020: EasyOne Pro LAB VOS Respiratory Analysis System Pulmonary Function Testing Device, Model 3100-1

Recall record Z-0597-2020 for NDDd Medical Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0597-2020
Company in FDA record
NDDd Medical Technologies, Inc.
FDA record creation date
Dec 3, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.

Source facts

Action
Customers were contacted via email or telephone between 8/1/17 and 4/13/19. The emails including the following information: We wanted to inform you that we are planning a preventive repair for some of our recently delivered EasyOne Pro/LAB V05 devices. The repair is related to an internal shielding component against electromagnetic disturbances. In some rare cases this could potentially lead to DLCO or MBW that cannot be performed. Although the probability of occurrence is very low, we want to absolutely ensure the quality of our devices and therefore we would like to make arrangements to update the shielding of this affected device. It is important to note, that all costs regarding the necessary repair will be covered by ndd, including all potential expenses. I would like to discuss handling of the repair and covering of the cost at the time of the repair. Of course, we already ensured that all newly delivered devices are not affected by the shielding issue. You can therefore order devices normally, there is no special action required. We are very sorry for the inconvenience this might be causing you as well as our shared customers and are committed to help as much as possible to ensure a smooth exchange of the affected component. Can you please contact me so we can make arrangements?
Initiated
2017-08-15
Root cause
Component change control
Terminated
2020-05-05
Distribution
The products were distributed to the following US states: KY, MO, NY, OH, PA
Recall number
Z-0597-2020
Recall status
Terminated
CDRH record ID
176934
FDA record created
2019-12-03
Recall event number
83992

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗