Reason stated in the source
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.
Source facts
- Initiated
- 2024-11-04
- Root cause
- Software design
- Distribution
- Worldwide distribution - US Nationwide and the countries of Austria, Bangladesh, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Malaysia, Netherlands, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom Great Britain, Vietnam.
- Recall number
- Z-0597-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210870
- FDA classified
- 2024-12-04
- Recall event number
- 95644