Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm
stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
Source facts
Action
Philips will send a customer notification to the users of the affected systems to alert them of the issue
and provide the actions that should be taken to prevent risks to patients.
Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and install the SSD spacer to resolve the issue.
Initiated
2025-10-27
Root cause
Radiation Control for Health and Safety Act
Distribution
U.S.
Recall number
Z-0597-2026
Recall status
Open, Classified
CDRH record ID
217505
FDA classified
2025-12-09
Recall event number
98052