The power manager subsystem may not initialize properly during start-up, which will cause the battery status monitoring and battery switching operations to not be fully functional.
Source facts
Action
Firm repaired the affected units beginning in November 2005, upon receipt of complaints of failure. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that they will be visited and their units upgraded as replacement components become available.
Initiated
2005-11-01
Root cause
Other
Terminated
2009-06-23
Distribution
Worldwide distribution ---- Nationwide and countries of Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Philippines, Republic of Georgia, Russia and Saudi Arabia.
Recall number
Z-0598-2007
Recall status
Terminated
CDRH record ID
48594
FDA classified
2007-06-30
Recall event number
36399