Z-0598-2025: Cartiva Synthetic Cartilage Implant (SCI) Catalog: CAR-06-US (6mm), CAR-08-US (8mm), CAR-10-US (10mm), CA

Recall record Z-0598-2025 for Cartiva, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0598-2025
Company in FDA record
Cartiva, Inc
FDA classification date
Dec 4, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Patients implanted with synthetic cartilage implant, may experience a higher-than expected occurrence rate of the following hazards: revision, removal, implant subsidence, displacement, pain, nerve damage or fragmentation.

Source facts

Action
On 10/31/24 recall notices were mailed or emailed to customers asking them to do the following: 1) Continue to follow patients treated with an impacted product for new or worsening symptoms of pain, difficulty walking, skin reactions, stiffness, swelling, or weakness of the big toe joint, consistent with your follow up protocols. Per Instructions for Use: the long-term effects of cartilage replacement are not known; and the clinical and medical status of each patient should be considered when treating patients. 2) To help minimize complications, reference the information in the Instructions for Use and the information included in this notification. Per standard practice, continue to discuss all potential risks identified and discuss the benefits and risks of all relevant treatment options for first metatarsophalangeal joint osteoarthritis with your patients. 3) Check your internal inventory to locate the products listed on the attached business reply form, remove them from their point of use, isolate/quarantine the unit(s), and return. 4) Share and maintain awareness of this communication in your practice with until all required actions have been completed within your facility. 5) If you have further distributed the affected product, please notify the applicable parties about this notice. You may copy and distribute this notification letter. 6) If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. 7) If you are a distributor, note that you are responsible for notifying your affected customers. 8) Complete and return the response form via email to fieldaction@stryker.com For questions or concerns, contact fieldaction@stryker.com.
Initiated
2024-10-31
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution in the states of CO, CT, FL, IL, MI, MS, NY, OH, RI, UT, VA, AL, AR, AZ, CA, IA, ID, IN, KS, KY, MA, MN, MO, MT, NC, NE, NM, NV, OK, OR, PA, TN, TX, WA, WI and the countries of Australia, Austria, Brazil, Canada, Chile, Cyprus, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Kuwait, Latvia, Lebanon, Netherlands, New Zealand, Panama, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Recall number
Z-0598-2025
Recall status
Open, Classified
CDRH record ID
210909
FDA classified
2024-12-04
Recall event number
95661

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗