Z-0599-06: Baxter Auto Syringe AS40 and AS40A Infusion Pump, product code 1M8565 (pumps with drug library option) an

Recall record Z-0599-06 for Baxter Healthcare Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0599-06
Company in FDA record
Baxter Healthcare Corp.
FDA classification date
Mar 7, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The occlusion detection and alarm may be delayed when the pumps are used with larger volume syringes at low flow rates.

Source facts

Action
Important Product Information letters dated 2/8/06 were sent to the AS40 customers via first class mail to the attention of the Director of Nursing. The accounts were informed of the delay in the detection of a downstream occlusion when the AS40 pump is used with large volume syringes at low flow rates. To minimize the occlusion detection times at low flow rates, Baxter recommended that the customers take the following actions: * Use the smallest possible syringe size for the volume of fluid to be delivered; * Select the low occlusion pressure setting to minimize the detection time; * Utilize infusion sets with thick-walled, small bore tubing to minimize the detection time; and * Always verify that the clamping device or stopcock is open in the fluid path prior to starting an infusion.
Initiated
2006-02-08
Root cause
Other
Terminated
2010-01-19
Distribution
Nationwide and Canada
Recall number
Z-0599-06
Recall status
Terminated
CDRH record ID
44130
FDA classified
2006-03-07
Recall event number
34579

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗