Skip to content

Public FDA record

Z-0599-2014: Asante Conset Infusion Set with 60cm tubing & 9mm cannula packaged in 5-packs,singles, or inside a Combin

Recall record Z-0599-2014 for Asante Solutions, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0599-2014
Company in FDA record
Asante Solutions, Inc.
FDA classification date
Dec 27, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Reports of the infusion tubing disconnecting from the connector, resulting in leakage of insulin from the infusion set and providing patient with less insulin than intended.

Source facts

Action
Asante sent urgent medical device recall notifications letters to distributors and customers on November 19, 2013 by US mail, return receipt requested.
Initiated
2013-11-19
Root cause
Device Design
Terminated
2014-04-24
Distribution
Nationwide Distribution.
Recall number
Z-0599-2014
Recall status
Terminated
CDRH record ID
124077
FDA classified
2013-12-27
Recall event number
66895

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗