Z-0599-2018: 3M Bair Hugger(M) Normothermia System, Sterile Full Access Underbody, Model 63700 The 3M Bair Hugger Te

Recall record Z-0599-2018 for 3M Company - Health Care Business. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0599-2018
Company in FDA record
3M Company - Health Care Business
FDA record creation date
Feb 12, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Adesign change to 3M(TM) Bair Hugger(TM) Full Access Underbody Warming Blanket, Models 63500 and 63700, can make these blankets more susceptible to unintended occlusion of airflow resulting in limited or partial inflation during use.

Source facts

Action
The firm, 3M, sent an "URGENT MEDICAL DEVICE RECALL" letter dated 1/19/2018 to its customers on 01/23/2018. The end user letter requested the following: "Actions Required: 1) Please immediately identify the affected product listed above, remove from your inventory and DO NOT USE. 2) Follow the instructions on the attached Customer Product Recall Acknowledgement Form, complete the form and return it to 3M via email at 3Mhealthcare@montage72.com. 3) Please dispose of all affected product per your institutional protocol and document disposal in the recall acknowledgment form." The distributor letter requested the following: "Actions Required: 1. Immediately discontinue distribution of the 3M(TM) Bair Hugger(TM) Full Access Underbody Warming Blanket, Models 63500 and 63700, with the referenced lot numbers above. 2. Determine if any stock of the 3M" Bair Hugger" Full Access Underbody Warming Blanket, Models 63500 and 63700, with the referenced lot numbers above remains in your inventory. 3. Complete the attached Distributor Product Recall Acknowledgement Form and return to 3M via email at 3Mhealthcare@montage72.com. It is important to notify 3M whether or not your facility has any inventory of this product. 4. Notify any of your customers who may have received the lots. Please use the attached customer notice and form provided in your communications to the customer. Customers should complete and return the enclosed form to 3M per the instructions on the form. Customers should not use the affected product and dispose per institutional protocol. 5. Dispose of any affected product in your inventory and document disposal in the recall acknowledgment form." If you have questions regarding this matter, please contact the 3M Health Care Customer Helpline at 1-800-228-3957, option #6, Monday through Friday, 7:30 a.m. 6:00 p.m. Central Daylight Time.
Initiated
2018-01-23
Root cause
Finished device change control
Terminated
2020-08-21
Distribution
Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, Chile, Costa Rica, Deutschland GMBH, Dominican, Ecuador, Hong Kong, Indonesia, Japan, Korea, Paraguay, Philippines, New Zealand, Shanghai, Thailand, and Uruguay.
Recall number
Z-0599-2018
Recall status
Terminated
CDRH record ID
161065
FDA record created
2018-02-12
Recall event number
79002

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗