Skip to content

Public FDA record

Z-0599-2019: AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 anal

Recall record Z-0599-2019 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0599-2019
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Dec 13, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.

Source facts

Action
On September 28, 2018, the firm, Tosoh Bioscience, notified consignees via an "URGENT: MEDICAL DEVICE RECALL" letter. The letter described the product, problem and actions to be taken. The letter informed customers that the AIA-900 and AIA -2000 analyzers had a potential issue that could result in slippage of the incubator unit within the system. Immediate Actions to be taken by the Customer/User " Thoroughly review the content of this letter. " Continue to use your Instrument. o Contact Tosoh Technical Support if mechanical errors occur. o The tension of the timing belt will be assessed and adjusted as needed at the next scheduled preventive maintenance visit. " Complete and return the attached Acknowledgement Form via Fax to: 1-605-636-8651; email: biorecallresponse@tosoh.com or mail to: Recall Coordinator, Tosoh Bioscience, Inc., 3600 Gantz Road, Grove City, OH 43123. " File this notice with your laboratory records and forward this information to others in your laboratory. If you have obtained unexpected test results, delays in test result reporting, or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. Should you have any questions regarding this medical device recall, please feel free to call (800) 248-6764 or email: bernadette.oconnell@tosoh.com Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Initiated
2018-09-28
Root cause
Process control
Terminated
2020-09-03
Distribution
US Nationwide distribution and International distribution to Colombia, Chile, Honduras, Cayman Islands, Panama, Costa Rica, Guatemala, Venezuela, Uruguay, Peru, and Ecuador.
Recall number
Z-0599-2019
Recall status
Terminated
CDRH record ID
167742
FDA record created
2018-12-13
Recall event number
81190

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗