Z-0601-05: Medtronic Maximo DR, model 7278 Single Chamber Implantable Cardioverter Defibrillator (ICD)devices are ti

Recall record Z-0601-05 for Medtronic Inc. Cardiac Rhythm Managment. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0601-05
Company in FDA record
Medtronic Inc. Cardiac Rhythm Managment
FDA classification date
Mar 11, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Medtronic marquis family of ICD and CRT-D devices having batteries manufactured prior to December 2003 that may experience rapid battery depletion due to a specific internal battery short mechanism. Once a short occurs, depletion can take place within a few hours to a few days, after which there is a complete loss of device function.

Source facts

Action
Medtronic Sales Representatives contacted affected physicians starting 02/10/05 to deliver an Important Patient Management Information. The letter recommended patient managment options including quarterly follow-ups, programming Patient Alert to 'On-High', provide magnet to patients to check device status and inform patients that should they expereicnce warmth in the area surrounding the ICD to seek prompt care.
Initiated
2005-02-09
Root cause
Other
Terminated
2006-12-28
Distribution
worldwide
Recall number
Z-0601-05
Recall status
Terminated
CDRH record ID
37337
FDA classified
2005-03-11
Recall event number
31142

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗