Reason stated in the source

Tosoh Bioscience has become aware of potential degradation of HbA1c control lot # 7055 before the 60-days product stability claim. If degradation of these controls occurs, P00 peaks may be seen and the % HbA1c control values may drift higher than the assigned range. This may result in delays when performing routine instrument quality control. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh has received two (2) complaints related to this issue with no serious injuries reported.

Source facts

Action
The firm sent notification to consignees on October 5, 2018, via Urgent Medical Device Recall letter. The letter informed consignees of the potential degradation of HbA1c control lot 7055 prior to the 60-days product stability claim. Immediate Actions to be taken by the Customer " Review the content of this letter with your laboratory staff. " Inform laboratory staff to avoid storing the control material in direct exposure to light. " Complete and return the attached Acknowledgement Form to Tosoh Bioscience, Inc., within 15-days of receiving this notification. " In the event of an early degradation of the control material, new controls should be used to verify the instrument performance prior to use. " Maintain this notification with your laboratory records and forward this information to others in your laboratory. " If you have obtained unexpected test results, delays in test result reporting, or received any complaints of illness or adverse events associated with the use of the analyzer, please contact Tosoh Technical Support 24 hours a day, seven days a week at (800) 248-6764. " Should you have any questions regarding this medical device recall, please feel free to contact Bernadette O'Connell at (800) 248-6764 or by email at bernadette.oconnell@tosoh.com. She will be available to answer any questions Monday - Friday, from 9:00 AM to 5:00 PM (PST).
Initiated
2018-10-05
Root cause
Device Design
Terminated
2020-06-16
Distribution
Distributed nationwide. Foreign distribution to Canada.
Recall number
Z-0601-2019
Recall status
Terminated
CDRH record ID
167736
FDA record created
2018-12-13
Recall event number
81188

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗