Z-0602-2014: Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise D

Recall record Z-0602-2014 for Integra LifeSciences Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0602-2014
Company in FDA record
Integra LifeSciences Corp.
FDA classification date
Dec 27, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.

Source facts

Action
Integra sent an Urgent Medical Device Correction letter and Correction Acknowledgement form dated November 26, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Our records indicate that you received one or more of the affected serial numbers that will require this firmware update. Integra asks that you do the following: 1. Review your inventory and verify that you have Camino Advanced Monitors with a serial number formatted as MExxxxxx04. If so, you can continue to use the CAM01 monitor(s), being aware that if you choose to use the optional SYS/DIAS mode, the values on the CAM01 display will be reversed. 2. Complete the attached form. If you do not have the product with the affected serial number in service at your facility, check the box: I do not have any product with a serial number of the form MExxxxxx04. 3. If you do have the affected serial number, check the box: I do have the product with a serial number of the form MExxxxxx04, and please indicate the serial numbers you have. 4. Complete the other information and return it by email or fax as indicated on the form. Complete the form and return it by email or fax as indicated on the form. Should you have any questions regarding these instructions, please contact Customer Service at 1-855-532-1723. We apologize for any inconvenience this may cause and thank you for your cooperation in this effort.
Initiated
2013-11-26
Root cause
Device Design
Terminated
2015-02-13
Distribution
Worldwide Distribution - USA ( Nationwide ) and the countries of : UK, Saudi Arabia, South Africa, Brunei, China, France, Poland, Turkey, Greece, Sweden, United Arab Emirates, Russia, Spain, Pakistan, Finland, Israel, Italy, Slovenia, Japan, Poland, Malaysia, Germany and Panama.
Recall number
Z-0602-2014
Recall status
Terminated
CDRH record ID
124692
FDA classified
2013-12-27
Recall event number
67094

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗