Z-0602-2019: AIA-900 Automated Immunoassay Analyzer performs three methods of immunoassay: an immunoenzymetric (IEMA)

Recall record Z-0602-2019 for Tosoh Bioscience Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0602-2019
Company in FDA record
Tosoh Bioscience Inc
FDA record creation date
Dec 13, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has become aware of a potential issue that may affect the mixing motor drive harness and mixing home detection sensor relay harness on your AIA-900 Analyzer. It is possible that the harnesses could rub against the Plarail chain during operation and result in premature wear. If observed, this issue will cause the following system error to be reported D.lane Mix home overrun error [4093] . If observed, this issue will cause the AIA-900 Analyzer not to function and the D.lane Mix home overrun error [4093] will be reported by the system. Replacement of the motor drive harness and mixing home detection sensor relay harness will mitigate the potential for premature wear and effectively prevent this issue from causing the system error. If the sensor relay harness fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh is aware of two complaints associated with the unexpected wearing of the harness. There have been no user or patient injuries reported that are related to this device malfunction.

Source facts

Action
On September 28, 2018, the firm notified customers via Urgent Medical Device Recall letter. Customers were informed of the potential issue. Immediate Actions to be taken by the Customer: Thoroughly review the content of this letter. " Identify if the serial number of your analyzer is listed in this notification. o If the serial number of your analyzer is listed: 1. At the next service call or scheduled PM, whichever occurs first, a Tosoh field service engineer will verify the motor drive harness and mixing home detection sensor relay harness are bundled together and placed into a double-coated cable or repair the harness bundle as applicable. 2. Complete and return the attached Acknowledgement Form by any of the listed methods. o If the serial number of your analyzer is not listed: 1. Tosoh records indicate that the harness bundle on your instrument(s) has already been replaced. Complete and return the attached Acknowledgement Form by any of the listed methods. " File this notice with your laboratory records and forward this information to other supervisors or managers in your laboratory to ensure that they are aware of the potential issue.
Initiated
2017-09-28
Root cause
Component design/selection
Terminated
2020-12-08
Distribution
Nationwide distribution. Foreign distribution to Chile, Colombia, Ecuador, Guatemala, Honduras, Panama, Peru, Puerto Rico, Uruguay, and Venezuela.
Recall number
Z-0602-2019
Recall status
Terminated
CDRH record ID
167650
FDA record created
2018-12-13
Recall event number
81172

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗