Open original FDA record ↗Recall record
- FDA identifier
- Z-0603-2016
- Company in FDA record
- Steris Corporation
- FDA record creation date
- Jan 6, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm discovered that information printed on the quick reference magnet is incorrect; specifically, Fahrenheit to Celsius temperature conversions and the default Warming Cabinet temperature.
Source facts
- Action
- The firm, Steris, sent an "URGENT VOLUNTARY FIELD CORRECTION NOTICE" letter dated 12/14/2015 to its customers. The letter described the product, problem and actions to be taken. The firm included an updated quick reference magnet for each affected unit with the letter. The customers were instructed to replace your existing magnet with the enclosed magnet. If you have questions regarding this matter, please contact Product Manager, at 440-392-7705 or STERIS Customer Service at 1-800-548-4873.
- Initiated
- 2015-11-20
- Root cause
- Process change control
- Terminated
- 2016-10-27
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Bahrain, Canada, China, Colombia, Costa Rica, France, Hong Kong, India, Indonesia, Israel, Kuwait, Lebanon, Mexico, Panama, Peru, Republic of Korea, Russian Federation, Saudi Arabia, Singapore, Spain, Taiwan, Thailand, Turkey, Uganda, United Arab Emirates, Viet Nam & Virgin Islands.
- Recall number
- Z-0603-2016
- Recall status
- Terminated
- CDRH record ID
- 142305
- FDA record created
- 2016-01-06
- Recall event number
- 72825