Z-0603-2019: HeartStart MRx, Rechargeable Li-Ion (lithium ion battery), REF: M3538A, which is a component of HeartStar

Recall record Z-0603-2019 for Philips North America LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0603-2019
Company in FDA record
Philips North America LLC
FDA record creation date
Dec 13, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected lithium-ion batteries may contain a defective component (Thermal Cut-off or TCO). Should the component fail, the battery will no longer charge or deliver power to the Monitor/Defibrillator, which may prevent operation of the device if it is not connected to AC or DC power. Should a second, unaffected battery be present in the Monitor/Defibrillator, the battery failure may go unnoticed.

Source facts

Action
On 11/23/2018 Urgent Medical Device Recall notices were mailed via USPS certified mail to customers. Customers were instructed to do the following. Inspect the batteries installed in your monitor/defibrillator to determine if they are affected by this recall. If any affected batteries are found, remove them from service immediately, along with any other affected batteries in your inventory. Further, customers were asked to complete and fax the Customer Reply Form to: +1.877.499.7223 or email to recall.response@philips.com. Your monitor/defibrillator can only be returned to service once a charged battery, unaffected by this recall, is installed. Firm will replace affected Rechargeable Lithium-Ion Batteries at no charge to the customer. Firm will contact you when replacement batteries are available. Customers with additional questions are encouraged to call 1-800-722-9377. Distributors were advised to send a copy of the Urgent Medical Device Recall notice and Customer Reply form to any customer to whom they have distributed affected batteries.
Initiated
2018-11-23
Root cause
Nonconforming Material/Component
Terminated
2020-09-29
Distribution
U.S.: AZ, WA, PA, UT, NY, CO, MD, AR, TX, MO, MI, FL, VA, MA, GA, TN, NV, OH, CA, NH, KY, KS, OR, WI, NC, IL, LA, NJ, WV, AL, WY, MS Foreign (OUS): United Kingdom, Korea Rep, Singapore, Australia, Taiwan, Sri Lanka, Brazil, France, Switzerland, Germany, Austria, Spain, Italy, Trinidad and Tobago, Estonia, Belgium, Portugal, Korea, Republic of, Argentina, India, Puerto Rico, Qatar, Netherlands, Mexico, South Africa, Ireland, Saudi Arabia, New Zealand, Bolivia, Costa Rica, Chile, Canada, Russian Federation, Slovenia
Recall number
Z-0603-2019
Recall status
Terminated
CDRH record ID
168731
FDA record created
2018-12-13
Recall event number
81443

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗