A problem can occur with the small light marker windows, used for the positioning laser and are integrated in the front cover of the SOMATOM Definition AS and Definition Flash systems, by either becoming loose or potentially dropping out. There is a risk that a person could touch rotating or electrical parts of the gantry if they were to reach through the window opening.
Source facts
Action
Siemens' sent a Customer Safety Advisory Notice letter dated October 19, 2016 to affected customers. The letter identified the affected product, problem, potential risks, how to avoid the potential risk related to this issue and actions to be taken.
Initiated
2016-10-19
Root cause
Nonconforming Material/Component
Terminated
2017-03-13
Distribution
US Nationwide Distribution
Recall number
Z-0604-2017
Recall status
Terminated
CDRH record ID
150783
FDA classified
2016-11-17
Recall event number
75555