Z-0604-2020: Eagle 3000 Sterilizer, small steam sterilizer. Product Usage: The Eagle 3000 Small Sterilizer is designed

Recall record Z-0604-2020 for Steris Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0604-2020
Company in FDA record
Steris Corporation
FDA record creation date
Dec 5, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for the incorrect pressure safety valve being included in the Preventive Maintenance (PM) kit for the 3000 Small Sterilizers.

Source facts

Action
On October 22, 2019 the firm distributed Urgent Medical Device Recall Communication letters by mail stating: URGENT DEVICE FIELD CORRECTION ATTN: MATERIALS MANAGEMENT OR STERILE PROCESSING DEPARTMENT RE: Pressure Safety Valve Kit to Eagle 3000 Small Sterilizers Dear Valued STERIS Customer: STERIS Corporation is voluntarily implementing a field correction for Eagle 3000 Small Sterilizers that may have had an incorrect pressure safety valve installed. Safety valve kits with incorrect valves were shipped between July 17, 2019 - August 28, 2019. Our records indicate that your facility was shipped one or more of the pressure safety valve kits included in this field correction. Description of the product - The Eagle 3000 Small Sterilizer is designed for sterilization of certain materials used in healthcare facilities, and are available in two configurations: " Gravity - designed for sterilization of non-porous, heat- and moisture stabile goods, and sterilization of items intended for immediate use. " Prevacuum - designed for fast, efficient sterilization of porous, heat- and moisture-stabile materials, in addition to the same sterilization capabilities as a Gravity sterilizer. Description of the problem - STERIS has identified that a small number of preventive maintenance kits for the Eagle 3000 Small Sterilizer were assembled with a I 00 psig safety valve instead of a 40 psig valve, potentially resulting in the JOO psig valves being installed on sterilizers in the field. Use of the incorrect valve could result in the improbable scenario of water and steam leaking from the unit's fittings, or accumulation of steam in the chamber. STERIS has not received any reports of adverse events related to the issue described above. STERIS Action - A STERIS Service Technician will contact your facility to inspect your unit(s), and repair if necessary. User Action - STERIS requests users inspect their affected unit(s) to verify which valve is installed. The valve ca
Initiated
2019-10-22
Root cause
Packaging
Terminated
2020-05-06
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL AR AZ CA CO DE FL GA HI IA ID IL IN KS LA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV, and country of Canada.
Recall number
Z-0604-2020
Recall status
Terminated
CDRH record ID
177539
FDA record created
2019-12-05
Recall event number
84185

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗