Open original FDA record ↗Recall record
- FDA identifier
- Z-0604-2024
- Company in FDA record
- Abbott Laboratories
- FDA classification date
- Dec 21, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The devices contain a dry natural rubber (latex) subcomponent, which are not labeled as containing dry natural rubber(latex).
Source facts
- Action
- The recalling firm issued the Product Correction letter dated 10/19/2023 via FedEx priority overnight express delivery in the U.S. and either in person or mail, telephone, email, or FAX to the OUS consignees. The product correction is for instruments distributed in the U.S. and Brazil only. The letter explains the issue, lists all products involved (some of which have different 806 report numbers and are being submitted by Division III) and that a precautionary label indicating the presence of dry natural rubber (latex) will be applied to the devices by the local Abbott representative. The Impact to User Safety is provided and actions to be taken by the customer, which included: Follow good laboratory practice by wearing impervious gloves and other personal protective equipment when working on the listed Abbott products; complete and return the Customer Reply Form; if the consignee has forwarded the product(s) listed to other laboratories, inform them of this Product Correction and provide a copy of the letter; and retain this letter for their laboratory records.
- Initiated
- 2023-10-19
- Root cause
- Under Investigation by firm
- Distribution
- Distribution was made to AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY, including Puerto Rico. There was also government and military distribution. Foreign distribution was made to Brazil.
- Recall number
- Z-0604-2024
- Recall status
- Open, Classified
- CDRH record ID
- 204057
- FDA classified
- 2023-12-21
- Recall event number
- 93349