Open original FDA record ↗Recall record
- FDA identifier
- Z-0605-03
- Company in FDA record
- Baxter Healthcare Corp.
- FDA classification date
- Mar 5, 2003
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Meridian hemodialysis instrument may reset when in shutdown mode causing the system to reset with potential for patient increased bioburden exposure.
Source facts
- Action
- Baxter issued a Safety Alert letter dated 11/15/02 to the direct accounts informing them of the potential for incomplete disinfection of the Meridian instrument during the Renalin/Formalin disinfection cycle due to electomagnetic interference (EMI). The accounts were informed that they will be receiving a radio frequency interference (RFI) filter kit and installation instructions to eliminate EMI by 2/28/03. The accounts were given instructions to follow to ensure that the Renalin/Formalin disinfection cycle has been completed prior to use of the machine.
- Initiated
- 2002-11-15
- Root cause
- Other
- Terminated
- 2003-09-19
- Distribution
- Nationwide, Mexico, China, South Korea, Hong Kong
- Recall number
- Z-0605-03
- Recall status
- Terminated
- CDRH record ID
- 25117
- FDA classified
- 2003-03-05
- Recall event number
- 25079