Open original FDA record ↗Recall record
- FDA identifier
- Z-0605-2014
- Company in FDA record
- Steris Corporation
- FDA classification date
- Dec 28, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.
Source facts
- Action
- On September 10, 2013, the firm sent Urgent Voluntary Recall Notices to their customers.
- Initiated
- 2013-09-10
- Root cause
- Use error
- Terminated
- 2014-02-14
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of AK, Al, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY, and the countries of BAHRAIN, CANADA, COLUMBIA, GERMANY, ITALY, MALAYSIA, PHILIPPINES, ROMANIA, SPAIN, and THE NETHERLANDS.
- Recall number
- Z-0605-2014
- Recall status
- Terminated
- CDRH record ID
- 124585
- FDA classified
- 2013-12-28
- Recall event number
- 67036