Z-0605-2018: Lyophilized Chemistry Calibrator Level 1 Product Usage: Beckman Coulter Chemistry Calibrators are inten

Recall record Z-0605-2018 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0605-2018
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Feb 13, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unpaired lots of AU lyophilized chemistry calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used.

Source facts

Action
On August 28, 2017 a letter was sent to all the customers stating the following: Attention Beckman Coulter Customer, Beckman Coulter is initiating a field action for the product listed above. This letter contains important information that needs your immediate attention. ISSUE: Unpaired lots of AU Lyophilized Chemistry Calibrator level 1 and 2 have been distributed. Calibrator level 1 and 2 must be paired when used for calibration and cannot be used in any other combination. Paired calibrator lots are identified by the last two digits of the lot number. Calibrator Level 1 REF DR0070-1, Lot 6101K51 Calibrator Level 2 REF DR0070-2, Lot 6102K61 Incorrect pair Calibrator Level 1 REF DR0070-1, Lot 6101K61 Calibrator Level 2 REF DR0070-2, Lot 6102K61 Correct pair IMPACT: If lot numbers are not correctly paired, calibration can be run using incorrect set points which may generate erroneous results. QC failures could occur if incorrect pairs of calibrators are used. Calibrator Level 1 REF DR0070-1, Lot 6101K51 Calibrator Level 2 REF DR0070-2, Lot 6102K61 Incorrect pair If the above incorrect pair is run with Level 1 lot 6101K51and Level 2 lot 6102K51 set points, erroneous results could be generated for the following assays: o Bicarbonate : maximum 11% bias low o Calcium Arsenazo : maximum 5% bias low o Calcium oCPC : maximum 5% bias low o Inorganic Phosphorous : 6.4% bias high o Magnesium: 3.2% bias low If the above incorrect pair is run with Level 1 lot 6101K61 and Level 2 lot 6102K61 set points, erroneous results could be generated for the following assays: o Bicarbonate : maximum 15.8% bias high o Direct Bilirubin OSR6x181: 8.7% bias low ACTION: Check the level 1 and level 2 calibrators to make sure you are using correctly paired lots for calibration. Ensure correct set points are being used for the lot. Set points are in the IFU for each lot and can be found in each kit, or on the Beckman Coulter websites
Initiated
2017-08-24
Root cause
Under Investigation by firm
Terminated
2020-08-28
Distribution
Worldwide Distribution in the states of AL, AK, AZ, AK, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IO, KS, KY, LA, MA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NY, NM, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VI, WA, WV, WI, and WY and the countries of Brazil, Canada, Costa Rica, El Salvador, Guam, Paraguay, and Peru
Recall number
Z-0605-2018
Recall status
Terminated
CDRH record ID
160143
FDA record created
2018-02-13
Recall event number
78699

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗