Skip to content

Public FDA record

Z-0605-2019: Tritanium Posterior Lumbar (PL) cage intervertebral body fusion implant. Intended for an open, less inva

Recall record Z-0605-2019 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0605-2019
Company in FDA record
Howmedica Osteonics Corp.
FDA record creation date
Dec 13, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The surgical technique is being updated to caution against misuse due to reports of cage fractures occurring intra-operatively and post-operatively.

Source facts

Action
Urgent Medical Device Correction letters were sent to customers on 11/28/18.
Initiated
2018-11-28
Root cause
Under Investigation by firm
Terminated
2020-05-28
Distribution
US Nationwide
Recall number
Z-0605-2019
Recall status
Terminated
CDRH record ID
169021
FDA record created
2018-12-13
Recall event number
81587

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗